7.2.11 SPECification Sub-system

SPECification
:C
:L
:H
[:PASS]
:TEST
:UNIT
SPECification:C/L/H
Description Set the voltage (CC, CR, CP mode) or current (CV mode)
Syntax SPECification:H
Parameters For valid value range refer to respective specification.
Example SPEC:H <NRf+>[suffix]
Query Syntax SPECification:H?
Query Example SPEC::H?
Return Parameters <NR2> [CC,CR,CP mode Unit=Voltage], [CV mode Unit=Current]
Return Example 4.75
SPECification[:PASS]?
Type All Channels
Description Request GO-NG result reference to all channel specification.
Query Syntax SPECification?
Query Example SPEC?
Return Parameters 0 | 1
Return Example 0 (NG), 1 (GO)
SPECification:TEST
Description Start or close the specification test.
Syntax SPECification:TEST <NR1 | CHAR>
Parameters 0 | OFF, 1 | ON
Example SPEC: TEST 1
Query Syntax SPECification:TEST?
Query Example SPEC:TEST?
Return Parameters 0 | 1
Return Example 1 (ON)
specification.
SPECification:L
SPECification:C
SPEC:L <NRf+>[suffix]
SPEC:C <NRf+>[suffix]
SPECification:L?
SPECification:C?
SPEC: TEST OFF
Return all channels GO-NG result.