7.2.10 RESistance Sub-system

RESistance
:L1
:L2
:RISE
:FALL
RESistance:L1/L2
Description Set the static resistance level for constant resistance mode.
Syntax RESistance:L1 <NRf+>[suffix]
Parameters For valid value range refer to respective specification.
Example RES:L1 20 OHM
Query Syntax RESistance:L1?
Return Parameters <NR2> [Unit=OHM]
Query Example RES:L1?
Return Example 10
RESistance:RISE/FALL
Description Set resistive slew rate of constant resistance.
Syntax RESistance:RISE <NRf+>[suffix]
Parameters For valid value range refer to respective specification.
Example RES:RISE 2.5
Query Syntax RESistance:RISE?
Return Parameters <NR2> [Unit=OHM]
Query Example RES:RISE?
RESistance:L2 <NRf+>[suffix]
RES:L2 10 OHM
RES:L1 MAX
RES:L2 MIN
RESistance:L2?
RESistance:L1? MAX
RESistance:L2? MIN
RESistance:FALL <NRf+>[suffix]
RES:FALL 1A/µS
RES:RISE MAX
RES:FALL MIN
RESistance:FALL?
RESistance:RISE? MAX
RESistance:FALL? MIN
Set constant resistance = 20 ohm for Load
L1.
Set constant resistance = 10 ohm for Load
L2
Set constant resistance = max.L1 value for
Load L1.
Set constant resistance = min. L2 value for
Load L2.
Return the set resistance value of Load L1
Set CR rise slew rate to 2.5A/µS.
Set CR fall slew rate to 1A/µS.
Set CR rise slew rate to the max.
programmable value.
Set CR fall slew rate to the min.
programmable value.
Return the CR rise slew rate.