Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
6.2.4 Command Format
The GPIB interface utilizes ASCII commands to allow remote control from a PC. The
command string is formed by [command+parameter]. The semicolon ";" is used to connect
multiple commands, ending with the End Code. The End Code is in one of the following
formats:
End Code
LF
CR + LF
EOI
LF + EOI
CR + LF + EOI
6.2.5 Panel Description
1.
Setting the Address
When "SYSTEM" appears in the title bar, press the ▲, ▼ keys to move the
●
highlighted cursor to [SYSTEM CONFIG] and press ENTER to go to the SYSTEM
CONFIG screen.
Press the ▲, ▼ keys again to move the highlighted cursor to [GPIB] and use the
●
[UP] and [DOWN] Function Keys to select the GPIB Address.
When the address has been set, press the [EXIT] Function Key to exit the menu.
●
2.
Remote & Panel Control
When the message box "Remote" is highlighted it indicates the tester is in the
●
remote control state.
●
In the remote control state, use the ENTER key on the front panel to switch the
tester to the panel control state.
In the remote control state, all keys are invalid except ENTER (switch to panel
●
control state) and STOP (reset the tester).
●
The GPIB LLO [Local Lockout] command can be used to lock the ENTER key.
6.3
RS232 Specification
6.3.1 Data Format
Baud Rate: 9600⁄19200⁄38400
Transmission Bit: 1 start bit + 8 data bits + 1 stop bit
6.3.2 Command Format
The RS232 interface utilizes ASCII commands to allow remote control from a PC. The
command string is formed by [command+parameter]. The semicolon ";" is used to connect
multiple commands, ending with the End Code. The End Code is in one of the following
formats:
6-2