High Power DC Electronic Load 63200 Series Operation & Programming Manual
7.2.8

POWer Sub-system

POWer
:L1
:L2
:RISE
:FALL
POWer:L1/L2
Description Set the Static Load Current of constant power mode.
Syntax POWer:L1 <NRf+>[suffix]
Parameters For valid value range refer to respective specification.
Example POW:L1 20
Query Syntax POWer:L1?
Return Parameters <NR2> [Unit=Watt]
Query Example POW:L1?
Return Example 3.12
POWer:RISE/FALL
Description Set current slew rate of constant power static mode.
Syntax POWer:RISE <NRf+> [suffix]
Parameters For valid value range refer to respective specification.
Example POW:RISE 2.5
Query Syntax POWer:RISE?
Return Parameters <NR2> [Unit=A/µS]
Query Example POW:RISE?
Return Example 2.5
7-20
POWer:L2 <NRf+>[suffix]
Set Constant Power = 20W for Static Load L1.
POW:L2 10
Set Constant Power = 10W for Static Load L2.
POW:L1 MAX
Set Constant Power = max. value for Static Load
L1.
POW:L2 MIN
Set Constant Power = min. value for Static Load
L2.
POWer:L2?
POWer:L1? MAX
POWer:L2? MIN
Return the set current value of the Static Load
L1.
POWer:FALL<NRf+> [suffix]
POW:FALL 1A/µS
POWer:FALL?
POWer:RISE? MAX
POWer:FALL? MIN
Set rise slew rate to 2.5A/µS for static load.
Set fall slew rate to 1A/µS for static load.
Return the rise slew rate of static load.