Chroma 63200 Series Operation & Programming Manual page 112

High power dc electronic load
Hide thumbs Also See for 63200 Series:
Table of Contents

Advertisement

High Power DC Electronic Load 63200 Series Operation & Programming Manual
STATus:QUEStionable:PTRansition/NTRansition
Description Programmable filters determine what type of transition (0-to-1 or
Syntax STATus:QUEStionable:PTRansition/NTRansition <NRf>
Parameters 0 ~ 65535
Example STAT:QUES:PTR 4
Query Syntax STATus:QUEStionable:PTRansition?
Return Parameters <NR1>
Query Example STAT:QUES:PTR?
Return Example 4
7-30
1-to-0) in the Condition register will set the corresponding bit of the
Event register.
STAT:QUES:NTR 4
STATus:QUEStionable:NTRansition?
Set OP(over power bit 2) to 0-to-1.
Set OP(over power bit 2) to 1-to-0.
Return the setting on the
QUEStionable Ptransition/Ntransition.