TroUBleshooTInG
symptom
Control lamps do not turn on when the
unit is switched on
No liquid flow when the foot pedal is
activated
No ultrasound oscillations or insuffi-
cient power
Water leak between the handpiece
cord and the handpiece
Water leak at the bottle or at the
peristaltic pump
86
solution
•
Check the connections of the power cable of the unit
•
Check the power supply of the dental office
•
Check the fuses and replace them only by the type of fuse indicated on the back of the unit
•
Send the unit to an approved EMS repair center
•
Turn the liquid flow rate ring to the maximum
•
Make sure that the bottle is properly installed
•
Check the connections between the unit, the handpiece cord and the handpiece
•
Check the connection of the foot pedal
•
Activate the "Cleaning" mode, disconnect the handpiece from the handpiece cord and check that liquid exits
the cord; if it is the case, the handpiece is plugged, send the handpiece to an approved EMS repair center
•
Unscrew the pump cover and check that the peristaltic pump functions
•
Send the unit to an approved EMS repair center
•
Make sure that the instrument is correctly tight
•
Check the wear of the instrument and replace it if necessary
•
Check the connections between the unit, the handpiece cord and the handpiece
•
Dry the connectors of the handpiece cord and of the handpiece
•
Check the connection of the foot pedal
•
Check the settings and the working mode
•
Check that when the handpiece is lifted from its support, the control lamps of the bottle holder are turned on; if
it is not the case, send the handpiece to an approved EMS repair center
•
Replace the handpiece cord and/or the handpiece
•
When one of the two handpieces is used, check that the other one is correctly seated on its support
•
Send the unit to an approved EMS repair center
•
Check the O-ring of the handpiece cord and replace if necessary
•
Send the handpiece cord to an approved EMS repair center
•
Send the unit to an approved EMS repair center
•
Check the O-ring of the bottle cap and replace if necessary
•
Replace the bottle
•
Replace the peristaltic pump
•
Send the unit to an approved EMS repair center