Reassembly - Dräger 8000 IC Operating Instructions Manual

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Cleaning / Disinfecting / Sterilizing

Reassembly

– Fan,
– seals,
– U-grommets,
– water hose,
– water bottle holder, or water feed system container.
Wash with detergent and rinse with clean water
or autoclave.
Bath disinfect parts. Allow exposure time for
disinfection as prescribed, then rinse with clear water
and dry;
or
sterilize at 120 °C (248 °F).
CAUTION !
When disinfecting the (reusable) skin temperature
probe, never immerse plug into the fluid.
Reassembly
Reassemble components in reverse order:
1 Fit trough.
2 Push fan wheel on axle.
3 Install base cover panel.
4 Install bed support brackets.
Check that the O-rings are correctly positioned on the
support brackets. If not, push rings all the way back
up or install new O-ring.
5 Lay bed on brackets. Watch for correct fit.
6 Fit mattress on bed.
Note:
If a ScaleTronix or other bed scale is used,
observe the operators manual of the scale for
cleaning and installation instructions.
6
1
5
2
4
3
4
3
3
4
1
6
Care
2
5
63