Using Air Temperature Control

Using Air Temperature Control
WARNING !
Always measure baby's core temperature at
regular intervals!
1 Push power switch in – until it engages = ON.
2 Green LED is lit.
Wait for self-test to complete. Incubator will then
switch to "air temperature control" mode.
NOTE: The air temperature control system is always
automatically activated at power-up.
3 This display will now alternate between measured
value for air temperature and the word SEt.
4 Flashing display of set value air temperature, default
value is 33.0 °C.
5 Press P or p key briefly to select this set value.
The display lights remain continuously lit.
Setting values in range between 28 °C and 37 °C
The set value can be changed in 0.1 °C steps.
6 Press p key – set value is decreased,
7 Press P key – set value is increased.
Press P or p keys until set value desired is
reached.
R R R R
8 Green LED
flashing – incubator is indicating that
heater is active.
°C LUFT/AIR
3
°C
>37 °C
4
Reset
Check
Control
°C LUFT/AIR
°C
>37 °C
5
5
Reset
Check
Control
°C LUFT/AIR
8
°C
>37 °C
6
7
Reset
Control
Operation