Dräger 8000 IC Operating Instructions Manual page 57

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Using Oxygen
Alarms
Switch off O
-Control
2
In the case of a pulled sensor module plug, a swung out
sensor module, or a defect sensor:
1 Red LED alarm will light and
2 red alarm-LED sensor will light, intermittend audible
alarm starts.
Then:
Immediately
connect sensor module,
swing back sensor module, or
change oxygen sensor.
The intermittend audible alarm can be silenced for 10
minutes:
3 Press G button,
g
4 LED
and
1 LED alarm will still light.
More about alarms in the chapter Troubleshooting.
Switch off O
-Control
2
– If any malfunction occures in the oxygen control
module, (Display Err),
– if there is a sensor alarm which can not be canceled,
– when even after decreasing the O
deviation to the setting higher that 5 % is,
– when oxygen therapy is finished.
5 Press ç button,
6 green module-LED Control does not light anymore.
Disconnect the O
supply from the incubator.
2
If the patient still requires an increased oxygen
concentration:
Refer to next chapter Oxygen administration without
the O
servo control module,
2
concentraion the
2
Vol.%
O
2
Vol.%O
2
>40%
Cal.
2
Control
21%
VOL.%O
2
6
Vol.%
O
2
Vol.%O
2
>40%
Cal.
Control
21%
5
Operation
1
4
3
Check
57