Care Disassembly - Dräger 8000 IC Operating Instructions Manual

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Care
Care
Clean and disinfect incubator after each patient, or at
least once a week. Perform all disinfection procedures
according to established hospital procedures as well as
to the following additional instructions.
WARNING !
Always follow established hospital procedures
for handling equipment contaminated with
bodily fluids.
Before any cleaning procedures, the incubator must first
be partially disassembled. For cleaning and disinfecting
of accessories, see their respective Operating
Instructions.
Disassembly
Switch off the device(s), pull power plug(s), and
turn off all pressurized gases that may be in use.
WARNING !
Always disconnect from power before
cleaning and disinfecting.
Remove all accessories.
Emptying water reservoir
Open and lower rear panel.
Remove all three water bottles.
Unscrew the two white plastice nuts.
1
Remove container by lifting it of the two bolts.
Drain water, observing established hospital
procedures.
Hold hose by cuff and detach from connector on
heater.
2
Open hand ports.
Remove seals from hand port openings and close
hand ports again.
60
Disassembly
Emptying water reservoir
2
1