What's What

Air Temperature Control Module

2.1
Key to switch on air temperature control (only with
skin temperature control option).
2.2
Unlocking key, extends range for set
temperatures
Yellow LED >37 °C,
2.3
is lit when a higher range of values has been set.
R R R R
2.4
Green LED
indicates warming-up phase.
2.5
Green LED Control,
stays lit while air temperature control is on.
2.6
Display for actual (measured) value of air
temperature.
2.7
Display for set value of air temperature;
left key: decreases set value
right key: increases set value
Red LED ±1.5 °C
2.8
flashes/is lit while the measured value of the air
temperature is deviating from the set value by more
than 1.5 °C.
2.9
Red LED Sensor
flashes when air temperature sensor is defective.
S S S S
2.10 Red LED
flashes in case of fan failure.
T T T T
2.11 Red LED
for overtemperature
Lights up when air temperature is or was higher
than 38 °C (40 °C during operation in extended
range) and after an inop-alarm.
2.12 Key to reset overtemperature alarm.
18
Air Temperature Control
°C LUFT/AIR
2.6
°C
°C
2.7
2.8
±1.5 °C
Reset
2.9
Sensor
2.10
2.11
2.12
Module