Operation
Using oxygen without the oxygen control
module
Monitor O
concentration. Always use calibrated
2
O
analyzer according to its Operating Instructions.
2
Supply oxygen from a medical gas pipeline system
or oxygen tank via an O
Adjust O
supply at flowmeter to required
2
concentration.
Guideline for flow settings:
Desired O
conc. 25 30 35
2
%
O
flow
2.0 4.5 6.5 9.0 11.5 14 16.5 19
2
L/min
58
flowmeter.
2
40 45
50 55 60