Dräger 8000 IC Operating Instructions Manual page 44

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Operation
Extending Temperature Range From
37 °C to 39 °C
WARNING !
Temperatures above 37 °C may only be used
on the order of a physician.
In this mode of operation, the baby´s
temperature must be monitored especially
carefully.
1 Press unlocking button 7
2 yellow indicator LED >37 °C lights up.
NOTE: This alters the temperature threshold for heater
cut-off in the event of overtemperature from
38 °C to 40 °C.
3 Keep P or p key pressed until the value desired is
displayed.
NOTE: Set value between 37.1 °C and 39 °C within the
next minute.
If extended range settings are not used within one minute
– control LED >37 °C will go off,
– blocking of extended temperature range will be
reactivated,
– the threshold for overtemperature will be reset to
38 °C.
Using a set value below 37 °C after use of the extended
range will also automatically cancel range extension.
NOTE: In case the incubator temperature is still above
38°C when leaving the extended range, the
overtemperature alarm (see next section Alarms)
would be triggered.
To avoid this alarm: First set value to 37.1 °C
and allow incubator to cool below 38 °C.
Then set desired value below 37 °C.
44
Extending Temperature Range
°C LUFT/AIR
2
°C
1
>37 °C
3
3
Reset
Check
Control