Technical Data
Range of set values
Air temperature
Skin temperature
Humidity
Measuring principles of sensors
Air temperature sensor
Skin temperature sensor
Noise level inside the canopy
Alarm Sound Level, outside
Alarm Sound Level, inside
Air filter
Skin temperature control
Sensor
Calibration tolerances
Sensor probe, reusable (disc-shaped)
Sensor probe, single use (miniature)
Temperature Variability
Temperature Uniformity
Classification
Manufacturer's certification of
electromagnetic compatiblity
Dimensions (height x width x depth)
Weight
76
28 °C to 39 °C
35 °C to 37 °C
35 to 85 % relative Humidity
NTC x 2
NTC x 2
56 dB (A) at 60 Hz;
according to IEC 601-2-19 more than 65 dB(A) at 3 m
according to IEC 601-2-19 less that 80 dB(A) over mattress
Particle class P 2, pursuant DIN 3181, 98 % efficiency
Use only sensors approved for use with
Dräger incubators
±0.1 °C
±0.1 °C
according to IEC 601-2-19 better than ±0.5 °C
according to IEC 601-2-19, range of ±0.8 °C
The incubator complies with DIN- VDE 0750, Part 1,
EN 60601-1, IEC 601-1 and IEC 601-2-19, CSA 22.2-125
m
Type B
The 8000 IC Incubator is certified to be free of radio
interference emissions pursuant guideline EN 55014:1987/A2
1990, and it complies with guideline 89/336/EEC regarding
immunity from electromagnetic interference.
52 - 60 x 38 x 27 inches, (1285 - 1480 x 960 x 680 mm)
197 lbs, (89kg)