Dräger 8000 IC Operating Instructions Manual page 45

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Alarms
Central Alarm
Alarms
Central Alarm
1 Red LED Alarm flashes and audible alarm will occure.
The intermittent audible alarm can be silenced for
10 minutes:
2 Press G key,
g
3 yellow LED
will be lit.
If there is a deviation of more than ±1.5 °C between set
value for air temperature and its measured value:
Red LED Alarm and
4 red LED ±1.5 °C flashes,
5 flashing display and intermittent audible alarm.
When measured value is again within ±1.5 °C of the set
temperature,
LED ±1.5 °C will go out and audible alarm will stop.
NOTE: When the incubator is switched on, the audible
alarm is automatically muted for 30 minutes
during warm-up.
4 red ±1.5 °C LED and
g
yellow
LED are lit.
If the air temperature is above 38 °C (or above 40 °C,
respectively, while using extended range):
T
6 Red LED
and
red LED Alarm are flashing,
7 flashing display and intermittent audible alarm.
The audible alarm can be silenced for 10 minutes.
Once the air temperature has fallen below the alarm
threshold again:
8 Press ƒ key to acknowledge alarm and to reset
heater.
°C LUFT/AIR
°C
>37 °C
Reset
Control
°C LUFT/AIR
5
°C
>37 °C
4
Reset
Control
°C LUFT/AIR
7
°C
>37 °C
Reset
Control
6
8
Operation
1
2
3
Check
Check
Check
45