Care
Installing hand port seals:
Insert seal marked with an "L" on its molded tag into
left opening
Insert seal marked with an "R"on its molded tag into
right opening
Position molded tag of seal directly on top of hinge.
Install with sealing lip (thin edge of profile) facing
outside.

Care for Humidification System

Allow water heater to run dry:
– to disinfect the heater
– to test that "water supply empty" alarm is working
properly.
Select humidity setting 85 %.
– Allow heater to run until "water supply empty" alarm
is triggered.
– Red LED H
O flashing,
2
– Display flashing accompanied by intermittent audible
alarm.
Switch off humidity control .
NOTE: When changing the water bottles or refilling the
reservoire, the water boiler will cool down due to
the fresh cold water. Expect the humidty in the
incubator to drop for 10 to 25 minutes depending
on the water temperature.

Before Reusing Incubator

For "Checking Readiness for Operation",
see page 34.
Operate incubator in standby,
or
switch off incubator and cover with dust cover, and
store ready for use.
64
Reassembly
Care for Humidity System
Before Reusing Incubator