Preparation; Before First Time Use; Front Flap; Quick Reference Flap - Dräger 8000 IC Operating Instructions Manual

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Before First Time Use

Front Flap

Quick Reference Flap

Preparation

Note: The incubator is delivered fully assembled.
Before First Time Use
Check that all packaging material has been removed.
Front Flap
To open the front flap
1 Squeeze catches with thumb and index finger and
open the front door.
Regular front flap: Guide flap slowly down until it rests
completly on the incubator base.
Scandinavian flap: Swing upper part of the flap up un-
til it rests completly on the hood.
To close the front flap
Press latch together with thumb and index finger and
press front flap against the hood. Once the front flap
is properly closed, the red index will no longer be
visible.
Scandinavian Hood, USA Standard
Quick Reference Flap
swing flap down to:
- switch on the incubator,
- change the set values
Then
swing flap up to eliminate the chance of unintended
changes to the settings.
To silence the intermittend audible alarm:
Press the with G labeled spot on the flap.
1
°C LUFT/AIR
°C FEUCHTE/HUMID.
°C HAUT/SKIN/PEAU/PIEL
°C LUFT/AIR
Control
Preparation
1
23

Advertisement

Table of Contents
loading

Table of Contents