Troubleshooting
Troubleshooting
Master Switch Module
Problem/Message
Red LED Inop lit.
Continuous audible alarm.
Red LED Inop lit.
Continous audible alarm,
with Err displayed in a module.
N N N N
Red LED
is lit.
Continuous audible alarm.
66
Likely Causes
General defect in electronics.
Defect in specific module:
Air temperature control module.
Skin temperature control module/
humidity control module.
No power.
Master Switch Module
Alarm
Inop.
Check
Remedy
Switch incubator off and on again.
If Inop message is not repeated:
press ƒ key and reset values.
If Inop message comes up again,
incubator is out of order .
Call DrägerService.
Switch off incubator.
Incubator out of order.
Call DrägerService.
Switch off respective module.
All other functions remain
operational.
Call DrägerService.
Check that power plug is connected
to line power.
Check that power is being supplied.
Inform biomedical engineering
and/or housekeeping.