Checking Hand Ports; Checking Front Door; Checking Air Filter; Testing Bed Tilt Mechanism - Dräger 8000 IC Operating Instructions Manual

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Checking hand ports, front
door and air filter

Testing bed tilt mechanism

Checking hand ports for proper closure
1 Grip rim of closed hand port and pull outwards
– it should not open.
In case hand port does not remain engaged:
Call DrägerService. Do not place a baby in the incu-
bator when a hand port does not close properly.
Always check all the hand ports.
Checking front door for proper closure
2 Squeeze catches together on both sides with thumb
and index finger – front door flaps open.
Close front door. Red marks on door catches should
no longer be visible indicating that the catches are
fully engaged. Check for correct closure by pulling
the door carefully fronwards.
If the front door does not engage properly:
Call DrägerService.

Checking air filter

Open back flap
3 Flip filter cover plate down.
If a filter is installed:
Remove filter and check installation date; write-on
label should be on edge of filter.
If filter is more than 2 months old:
Replace with a new filter and write installation date on
filter label.
Press filter firmly into the seal.
Make sure that the direction of flow through filter is
correct. Arrow on filter indicating flow direction must
point into incubator.
Testing bed tilt mechanism
4 Raise left side of the mattress to maximal height.
5 Press down on raised bed with hand – bed must not
drop down.
If bed drops down:
Call DrägerService.
Repeate the same procedure for the right side.
Checking Readiness for
1
1
2
3
5
4
Operation
1
2
35