Troubleshooting

Skin Temperature Control Module

Problem/Message
Red LED ±1.0 °C flashing.
Display of measured value flashing.
Intermittent audible alarm.
Red LED Sensor lit.
Three dashes flashing in center of
display for measured values.
Intermittent audible alarm.
Err flashing in display for measured
values.
Red LED Inop. lit in main module.
Continuous audible alarm.
Err continously lit in display of
measured values.
Reference temperature of 36 °C
is outside tolerance of ±0.1 °C
when 6 key is pressed.
68
Likely Causes
Skin temperature deviating from
set value by more or equal
than ±1.0 °C.
Skin temperature sensor not
properly connected.
Defective skin temperature sensor.
Defect in skin temperature control
module while in operation.
Defect in skin temperature control
module (while not in operation);
skin temperature control switched off. Call DrägerService.
Temperature measuring module
not operating accurately.
Skin Temperature Control
°C HAUT/SKIN/PEAU/PIEL
Control
°C
°C
± 0,5 °C
Check
15.22
Control
36 °C
87.06
±0,1
Sensor
Remedy
Below set value:
Check that sensor is properly
attached to patient.
Above set value:
Measure core temperature of
patient and inform physician re-
sponsible for patient immediately.
Check connection and correct
if necessary.
Replace sensor.
Switch to air temperature control.
Call DrägerService.
Other modules continue to operate
properly.
Call DrägerService.