Checking LEDs, displays and
alarm sound
Checking LEDs, displays and alarm sound
1 Press k button:
For about 2 seconds – all LEDs are lit (except the
power failure LED), the digital displays show 88.8 and
an audible alarm is generated.
Thereafter, displays and LEDs go dark and audible
alarm stops.
After approximately another 2 seconds, the original
displays for measured and set values reappear.
NOTE: This check can also be carried out during
operation and does not alter any of the settings.
Perform check at least once daily.
If there is a defect:
Call DrägerService.
WARNING !
The incubator is ready for operation only when
all checks have been performed successfully.
WARNING !
Do not use cellular or cordless phones within
33 feet (10 m ) of the equipment.
Wireless phones may cause malfunction in
electromedical equipment.
CAUTION !
Never block or obstruckt the air vents in the
incubator base-element, e.g. with blankets.
This may cause burns or cooling of the baby.
Checking Readiness for
°C LUFT/AIR
°C
>37 °C
Reset
Check
Control
Operation
1
5
37