Operation

Starting Operation

Wait for incubator warm-up
Wait for about 35 minutes before using incubator.
Heat incubator in "air temperature control" operating
mode.
In order to keep incubator ready for use without delay:
Operate incubator in standby:
Set air temperature between 32 °C and 36 °C
and, set humidity to 0.

Water Supply

Before placing baby in the incubator:
1 Open and lower rear panel.
2 Remove empty water bottles from receptacle and
replace with three full bottles.
Lift rear panel and close it.
– If your incubator 8000 IC is equiped with the
automatic water feed system change the attached
water bag.
NOTE: If an automatic water feed system is used,
observe the operators manual of the automatic
water feed system for operating and installation
instructions.
WARNING !
Use only pure, distilled or demineralized
water. Do not add bactericidal agents.
Adjust working height
– Change incubator height as needed and described in
the previous chapter.

Placing a baby in the incubator

Open front door (top and bottom part) and
pull bed out.
Put baby on mattress and slide bed back in position.
Close the front door and check that all latches are
properly engaged. Check hand port doors for propper
closure.
Keep bed horizontal or tilt, as required.
42
Starting Operation
Water Supply
Placing a Baby in the Incubator