Using Oxygen

Calibration of Oxygen Sensor
Using Oxygen
WARNING !
Oxygen Concentration
The atmosphere inside the incubator should
only be enriched with oxygen by or on the
order of a physician or respiratory therapist.
Always monitor oxygen concentration!
Connect the incubator with a high pressure oxygen
hose to the oxygen outlet of the pipeline system.
1 Press ç button,
2 Green module-LED Control iluminates.
3 Yellow LED Cal. and
4 Display CAL flashes
5 The set value of »21« flashes in the display
Calibrate O
sensor immediately, otherwise sensor
2
alarm.
Calibration of oxygen sensors
Calibrate oxygen sensor when starting with oxygen
control and every 24 hours if used continuously.
Swing sensor module all the way out from the
incubator hood. Do not use force!
Within one minute, otherwise sensor alarm:
6 Press 2 button,
7 yellow LED Cal. illuminates.
8 During calibration the display flashes alternately
---
between CAL and "
".
WARNING !
It is absolutely essential that elevated oxygen
concentrations are selected on the basis of
arterially measured oxygen partial pressure in
the blood of the baby. This is the only way of
minimizing the risk of both hyperoxemia,
which might cause, above all, retrolenta
fibroplasia, and hypoxemia which might
contribute to intraventricular hemorrhage and
damage to the baby's brain.
VOL.%O
2
4
Vol.%
O
2
Vol.%O
2
5
Cal.
21%
VOL.%O
2
8
Vol.%
O
2
Vol.%O
2
Cal.
21%
6
Operation
2
3
>40%
Control
1
7
>40%
Control
55