Dräger 8000 IC Operating Instructions Manual page 48

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Operation
NOTE: When a skin temperature sensor is connected
while "air temperature control" mode is active,
the measured skin temperature is still displayed.
However, it does not affect the incubator
temperature, which is controled only by the set
air temperature.
When the temperature falls outside the measuring range
of 30 °C to 42 °C:
3 dashes in the bottom of the display =
temperature below 30 °C
3 dashes in the top of the display =
temperature above 42 °C
Activating skin temperature control
NOTE: Allow at least 5 minutes for the skin temperature
sensor to reach the baby's temperature. Do not
switch to skin control when the temperature is not
constant.
When value displayed remains constant,
1 Press ç key,
2 green LED Control lights up,
skin temperature control is now active.
3 The large display alternates between the measured
value of skin temperature and the word SEt.
4 The small display for skin temperature set values is
flashing.
5 The measured value for air temperature continues to
be displayed.
48
Using Skin Temperature Control
Activating skin temperature
°C HAUT/SKIN/PEAU/PIEL
°C
Check
Control
36 °C
±0,1
°C
°C
°C HAUT/SKIN/PEAU/PIEL
2
3
5
°C
4
Check
36 °C
±0,1
1
control
°C LUFT/AIR
°C
>37 °C
Reset
Control
°C LUFT/AIR
°C
>37 °C
Reset
Control

Advertisement