Dräger 8000 IC Operating Instructions Manual page 61

Hide thumbs Also See for 8000 IC:
Table of Contents

Advertisement

Disassembly
If skin temperature control option has been used:
Disconnect skin temperature sensor.
Open front door.
Remove skin temperature sensor from inside
incubator.
1
Remove all silicone U-grommets.
Tilt incubator hood back as far as it will go. Do not
use force or push the hood further back.
2
Remove mattress from bed.
3
Lift off bed. Undo screw on ventilator circuit
support arm and remove flexible support arm.
4
Pull bed support brackets up and out.
5
Remove base cover panel.
WARNING !
Danger of burn injury
The exposed heater surface will still be very
hot after operation. It may take up to one hour
for the temperature to drop to 70 °C (158 °F)
with the incubator closed.
6 Remove fan.
7 Remove trough.
3
1
4
2
1
3
2
3
4
3
4
5
4
7
6
Care
5
6
61