Alarms
Alarms
In case of deviations greater or equal than ±1.0 °C
between set and measured value of skin temperature:
1 Red LED Alarm and
2 Red LED ±1.0 °C flashing,
3 display flashing accompanied by intermittent audible
alarm.
The audible alarm can be silenced for 10 minutes:
4 Press G button.
g
5 Yellow LED
lights up.
When measured value is again within less ±1.0 °C of the
set value:
LED ±1.0 °C and
1 LED Alarm goes out, audible alarm stops.
g
5 Yellow LED
goes out.
If the sensor plug is disconnected or sensor is defective:
6 Display of 3 dashes in center of display.
After 15 seconds:
intermitted audible alarm starts
6 Display of 3 flashing dashes in center of display.
7 Red LED Alarm flashing.
8 Red LED Sensor flashing.
In this case:
Connect sensor without delay or replace sensor.
The audible alarm can be silenced for 10 minutes:
9 Press G button.
g
10 Yellow LED
lights up.
8 Red LED Alarm lights up.