10. Maintenance

Replacing the window foil
The foils cover the foil frame. After removing of the protective
grid (2 ea. screws), the outer foil frame (single mylar foil) is with-
in reach. After dismantling the cover sheet and the pressure frame
(10 ea. screws), the outer foil frame can be easily taken out and
changed.
In order to change the second foil frame, the sealing frame has to
be removed. After removal, the defective foil frame has to be
carefully unfixed from the scintillator.
RadEye AB100
DB-070 E
10-1