9. Trouble Shooting

RadEye AB100 doesn't power on

9.1

Press "On" for at least one second

9.1.1
Keep the button "" pressed until the LED and the sound is ac-
tive. If the button is pressed only shortly, then the start up routine
is not activated. The LED may flash then only very shortly.
9.1.2

Check battery

If the RadEye does not respond to a long press (> 1s) of the ""
button, then please use the following step by step procedure:
A) Remove the current batteries
B) Check the battery contacts
C) Check the required orientation of the batteries according to the
printed "+" symbols in the battery compartment.
D) Insert two -"C" batteries with sufficient capacity: Rechargea-
ble batteries should be completely loaded, non rechargeable bat-
teries should supply together at least 2.3 Volt under load.
E) Press the "" button for more than one second (see also 2.3)
RadEye AB100
Pi/Ff 04.05.2021
DB-070 E
9-1