6. Functional test

When the instrument is switched on, it will show zero count rate
or activity until the first detector pulse occurs.
The radiation meter performs continuous self-check routines. A
timeout for the detector can be configured via the PC-software. A
complete failure of the detector during operation will be indicated
after the timeout on the LCD and be announced by the beeper.
The same applies to the battery voltage.
6.1
Functional test
To carry out a simple test, shortly press any key. A short audible
pulse has to be released and the LC display is illuminated for
some seconds.
The heart symbol next to the battery indicator must be "beating".
This indicates that the cyclic tasks as calculating measurement
values and checking for alarm thresholds are active.
RadEye AB100
DB-070 E
6-1
Pi/Ff 04.05.2021