RadEye lost date and time settings

The date and time information is lost, each time the batteries are
removed or below the required voltage for operation. The sim-
plest way to program the date and time is to use the "synchro-
nize" function of the optional RadEye.EXE software. Please refer
to chapter 7.3.2. Else the current date and time can be set also
through the RadEye's menu "Settings" / "Set Date / Time"

RadEye's battery status is lower than expected

9.7
Please check if the correct battery type is selected under "set-
tings" / "Batt. type" and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batteries
provide 1.5V.
9.8

History data shows the wrong time and date

See chapter 7.3.2.