Accumulated Counts; Background Measurement - Thermo Scientific RadEye AB100 Operating Instructions Manual

Multi-purpose survey meter
Table of Contents

Advertisement

R
R
=
M
F
t
t
M
0
F:
absolute error in cps
R
:
measuring value in cps
M
t
:
Measuring time in s
M
R
:
Background count rate *)
0
t
:
Background measuring time *)
0
*) Background count rate and background measuring time are
only used in combination with "Scaler netto"
3.3.2

Accumulated counts

With accumulated counts selected via radeye.exe the RadEye
AB100 shows the collected counts within the measurement time
in place of the mean value. This option is only available in com-
bination with measurement unit "cps" and "cpm".
3.4

Background measurement

To enter background measurement select in submenu "Back-
ground" "Start Bgr.meas".
The background measurement starts with the parameter set in
submenu option „Set parameter". To start the measurement, press
key. While the measurement is in progress "Exit" is
"Info/▲"
the
blinking.
RadEye AB100
Pi/Ff 04.05.2021
0
DB-070 E