values according to IATA, respectively NRC (USA) or
StrSchV(Germany) in respect to the scope of dangerous goods
definition and transport regulations of radioactive material.
6.5

Low Battery warning

If the battery voltage is below the configured threshold (see chap-
ter 4.12) the following warning appears:
The beeper generates a melody every 32s. This acoustic warning
can be suppressed by pressing the alarm acknowledge key. After
8h this warning comes up again. The battery needs to be changed.
However, the RadEye can still be operated for several hours.
If the battery voltage falls below 2V, the battery symbol starts
flashing. An acoustic warning is generated every 8s and can't be
acknowledged. The battery needs to be changed as soon as possi-
ble. Dead batteries may leak. Remove them from the RadEye to
prevent corrosion.
Battery life is significantly decreased at low temperatures. Cells
which give a mid-scale voltage reading in a warm building may
fail rapidly if the instrument is then operated outdoors in winter.
6-4
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021