2.2

Inserting the battery

The two C-Alkaline cells as delivered with the instrument
allow about 1200 h of normal operation.
C size rechargeable batteries can be used as well.
• Switch off the measuring instrument.
• Open the cover of the battery compartment.
• Exchange the batteries according to the shown polarity.
• Close the compartment cover.
• Switch on the unit again. (see chapter 2.3)
The instrument continues working in the operating mode set
last (see chapter 3.2).
The measured values in the history memory remain stored.
The real time clock for time stamp of history values and log-
book continue operation, if batteries are exchanged without
delays. If Real Time Clock is set, actual time and date is dis-
played for 3 s (see chapter 2.3).
After selection of the counter tube the RadEye AB100 first
display 0.00 cps. After the first detector pulse the measure-
ment will be started.
If correct timing information of logbook and history is re-
quired, the Real Time Clock should be set by means of the PC-
program.