When the measurement is complete the values of the counter
tubes are read and shown graphically in the form of a chart and
also as a table. Subsequently the high voltage is increased by the
step size entered. After a stabilization time of 2 s has elapsed the
next measurement will start. This procedure is repeated until the
end voltage entered has been reached.
While the plateau measurement is being carried out all of the
buttons and menu functions are locked except for the stop button.
8.1.4
Printing the plateau
The "Print" button only becomes active after loading or measur-
ing a plateau. Once the button has been pressed, the plateau is
printed using the Windows printer that is currently installed.
8.1.5
Saving a measured plateau
A measured plateau is automatically saved after the last meas-
urement has been completed. Both the measured values and the
parameters are saved to a monthly file. The name of the file is
RadEyeAB100_yymm.plt, where mm stands for the month, and
yy for the year (Example: The monthly file of January 2010 is
named RadEyeAB100 _1001.plt).
8.1.6
Displaying saved plateau data
Use the "File" "Open" menu point to open a dialog window
where any desired monthly file "RadEyeAB100_*.plt" can be
selected and reviewed.
RadEye AB100
Pi/Ff 04.05.2021
DB-070 E
8-7