Window: Here, the right windows configuration can be set.
o no window configuration (Rate 1, Rate 3)
o Beta window (Rate 1 – Rate 2), Rate 3
With "Plateau measurement" the user is able to perform a plateau
measurement (see chapter 8)
Clicking on "Extended" the user may configure some more op-
tions:
Dead time: In these fields a dead time correction factor for each
counter can be edited.
The dead time corrected count rate is calculated as follows:
CR:
CRraw:
DT:
The denominator "1- CRraw*DT" is limited between 0.1 and 1.
7-12
=
CR
1−
Count rate with dead time correction
Raw count rate without dead time correction
dead time in seconds
DB-070 E
CRraw
CRraw
*
DT
RadEye AB100