Surface Contamination (Bq; Dps; Dpm; Bq/Cm ) - Thermo Scientific RadEye AB100 Operating Instructions Manual

Multi-purpose survey meter
Table of Contents

Advertisement

2
5.2
Surface Contamination (Bq; dps; dpm; Bq/cm
)
Up to 16 nuclides for each counter tube together with their related
calibration factors can be stored in the RadEye AB100. The name
of isotopes, the number of available nuclides as well as their 4 Pi
efficiency can be edited in the PC-software.
The user of the RadEye AB100 can select the required isotope in
the menu <Nuclide>. This selected nuclide (e.g. Co-60) is dis-
played in the LCD (eventually alternating to the subtracted back-
ground activity in case of a net measurement).
If a no nuclide is configured, the RadEye AB100 shows "###".
If a nuclide is configured only with α- or β-value, the measure-
ment without value creates also the RadEye AB100 display
"###".
For example: β efficiency for Sr-90 is set to 45% and α efficiency
is set to 0%. The display for α contamination is "### Bq".
5-2
DB-070 E
RadEye AB100
Pi/Ff 04.05.2021