GE Healthcare
Immunity Test
NOTE:
* Field strengths from fixed transmitters, such as base stations for cellular telephones
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be estimated accurately. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be performed. If the measured
field strength exceeds the RF compliance level above, observe the Innova system to
verify normal operation in each use location. If abnormal performance is observed,
additional measures may be necessary, such as re–orienting or relocating the Innova
system
**Over the frequency range 150 kHz to 2.5GHz, the recommended field strengths is here
less than 0.3V/m.
The Recommended Separation Distances are listed in
These are guidelines. Actual conditions may vary.
2.3.3.3 Recommended Separation Distances for Portable and Mobile RF Communications Equipment
IEC 60601–1–2
Table 3-10:
Frequency of Transmitter
Equation
Rated Power of Trans‐
mitter (watts)
10 mW
100 mW
1
10
100
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE:
These are guidelines. Actual conditions may vary.
62
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
IEC 60601-1-2 Test Level
150 KHz to 26 MHz
d = [3.5 / V 2 ]√P
Distance (meters)
1.2
3.8
12
38
120
Compliance Level
Note: P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the rec‐
ommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,* are less than the compliance
level in each frequency range.**
Interference may occur in the vicinity of equipment marked
with the following symbol:
26 MHz to 80 MHz
d = [3.5 / V 2 ]√P
Distance (meters)
1.2
3.8
12
38
120
2 Environmental Requirements/Limitations
Direction 5177951-4-1EN, Revision 1
Electromagnetic Environment
Table
3-10.
80 MHz to 800 MHz
d = [3.5 / E 2 ]√P
Distance (meters)
1.2
2.3
3.8
7.3
12
23 (*)
38
73
120
230
800 MHz to 2.5 GHz
d = [3.5 / E 4 ]√P
Distance (meters)