Initial System Power-On

Review the previous sections in this chapter and verify that all site installation requirements have
been met. The laser system should only be operated in an area of restricted access. Post a laser
warning sign on each access door.
1. Connect the foot switch and remote interlock connector to the laser system. (See Figure 3,
page 1-6.) Find the remote interlock connector supplied separately in the bag that comes with
the unit.
2. Connect the laser device to a suitable AC electrical outlet.
WARNING!
ply MAINS with protective earth.
3. Ensure that the emergency stop button isn't engaged. (See Figure 1, page 1-5.)
4. Turn the MAINS switch on. (See Figure 3, page 1-6.)
NOTE:
Position the system so both the MAINS switch and power connection to the electrical
outlet can be easily reached.
5. Turn the key switch on. (See Figure 2, page 1-6.)
6. Lift the laser aperture cover and connect a 550 µm holmium laser fiber to the laser. (See Figure
1, page 1-5.)
NOTE:
Do not over-tighten the laser fiber. The connector only needs to be finger tightened
for proper operation.
7. After a few seconds, the information bar on the control panel screen will display "STANDBY. "
(See Figure 4.)
8. Press the ready button. (See Figure 6, page 1-32.) The "READY" progress bar will take about
two seconds to fill.
Cook Medical © 2019
To avoid the risk of electric shock, connect this laser system only to a sup-
Figure 4
Return to Table of Contents
Return to Table of Contents
Chapter 4
Installation
1-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rhapsody hl-30aRhapsody hl-30bG23663G23071

Table of Contents