Laser Parameters - COOK Medical Rhapsody H-30 User Manual

Hide thumbs Also See for Rhapsody H-30:
Table of Contents

Advertisement

Available languages

Available languages

Soft Tissue
Use higher pulse energy to create larger ablation volumes than lower pulse energy.
NOTE:
Higher pulse energy will also increase bleeding.
The information in the following sections provides further indications, contraindications,
warnings, and precautions for the Rhapsody H-30 Holmium Laser System for specific specialties as
appropriate.

Laser Parameters

The following tables display available laser parameters; possible combination cells
are marked with the average power (power = energy x frequency) in watts for all available
settings.
With a 150 µm fiber, the maximum power is 6 W.
With a 200 µm fiber, the maximum power is 10 W.
With a 273 µm fiber, the maximum power is 15 W.
With a 365 µm fiber, the maximum power is 20 W.
With a 550 µm fiber, the maximum power is 30 W.
With a 940 µm fiber, the maximum power is 30 W.
Combinations in Short pulse width mode:
Watts
5 Hz
0.5 J
2.5
0.6 J
3.0
0.7 J
3.5
0.8 J
4.0
1.0 J
5.0
1.2 J
6.0
1.5 J
7.5
2.0 J
10.0
2.5 J
12.5
3.0 J
15.0
3.5 J
17.5
Combinations in Long pulse width mode:
Watts
5 Hz
0.5 J
2.5
0.6 J
3.0
0.7 J
3.5
Cook Medical © 2019
8 Hz
10 Hz
12 Hz
4.0
5.0
4.8
6.0
5.6
7.0
6.4
8.0
8.0
10.0
12.0
9.6
12.0
14.4
12.0
15.0
18.0
16.0
20.0
24.0
20.0
25.0
30.0
24.0
30.0
8 Hz
10 Hz
12 Hz
4.0
5.0
4.8
6.0
5.6
7.0
Return to Table of Contents
Return to Table of Contents
15 Hz
20 Hz
6.0
7.5
10.0
7.2
9.0
8.4
10.5
9.6
12.0
15.0
18.0
22.5
15 Hz
20 Hz
6.0
7.5
10.0
7.2
9.0
8.4
10.5
Chapter 5
Operation
1-36