Pedal not connected: The foot switch connector is not connected or is connected incorrectly. This error warning is
displayed when the system is in Ready mode. Connect foot switch.
Energy < 80%: This warning appears when the output energy is less than 80% of the set value. Call for service.
Energy > 120%: This warning appears when the output energy is greater than 120% of the set value. Call for service.
Energy > 200%: This warning appears when the output energy is greater than 200% of the set value. Call for
service.
Maintenance due: This warning appears when preventative maintenance is due. Call for service.
MPS charge error: This error indicates the main power supply is not functioning properly. Call for service.
MPS over temp: This error indicates the main power supply is not functioning properly. Call for service.
MPS overload: This error indicates the main power supply is not functioning properly. Call for service.
Cook Medical © 2019
Return to Table of Contents
Chapter 5
Operation
1-38