Troubleshooting

Trouble
Key switch does nothing
No correspondence between
displayed energy and effects on
skin
Error Messages
Error Message
Low coolant flow
Over temperature
Shutter failure
Simmer failure
No fiber attached
Remote open
Pedal not connected
No energy
Energy < 80%*
Energy > 120%*
Energy > 200%
Cook Medical © 2019
Possible Cause
Power cord is not connected
Main line breaker is off
Emergency stop button is
pressed
Damaged fiber
Damaged blast shield
Possible Cause
Not enough water
Room temperature is too high
Airflow is obstructed
Cooling system component
failure
Component failure
Flash lamp not functional
Incorrect installation of
the fiber
Wrong fiber is installed
The remote interlock is not
connected or is opened
The remote interlock is damaged
The foot switch is not connected
The foot switch is damaged
Optical component damage
Optical component damage
Incorrect energy calibration
Incorrect energy calibration
Return to Table of Contents
Return to Table of Contents
Maintenance
Solution
Connect the power cord
Check the MAINS breaker
Release the emergency stop
button
Change the fiber
Change the blast shield
Solution
Fill the water reservoir
Lower the room temperature
Ensure nothing is obstructing the
fan
Call for service
Call for service
Call for service
Check the proper connection of
the fiber
Install suitable Rhapsody H-30
Holmium Laser System fiber
Check the remote interlock
connection
Call for service
Connect the foot switch properly
Call for service
Call for service
If the energy error persists, call
for service*
If the energy error persists, call
for service*
Call for service
Chapter 6