COOK Medical Hemospray Quick Reference Manual

COOK Medical Hemospray Quick Reference Manual

Endoscopic hemostat

Advertisement

Quick Links

Hemospray
E N D O S C O P I C H E M O S TAT
Prep 1
Remove device from package and attach
catheter to handle, ensuring connection
is secure.
Step 1
Before inserting catheter into accessory
channel, identify bleeding site, remove as
much blood as possible, then flush accessory
channel with air.
M E D I C A L
®
Prep 2
Activate CO
cartridge by turning
2
knob until it stops.
Step 2
Slowly advance catheter through accessory
channel in short increments until catheter tip
is visualized endoscopically. Precaution: To
avoid catheter occlusion, do not place catheter
directly in contact with blood and/or mucosa,
including any pooled blood and do not aspirate
blood while catheter is in accessory channel. To
ensure proper visibility, catheter tip should be
1-2 cm away from bleeding site at all times.
Q U I C K R E F E R E N C E G U I D E
red
activation
Step 3
To allow powder deployment, turn
open position. Note: Device is now active and
ready for use. Do not press trigger button until
powder deployment is desired.
red
valve to

Advertisement

loading

Summary of Contents for COOK Medical Hemospray

  • Page 1 Q U I C K R E F E R E N C E G U I D E Hemospray ® E N D O S C O P I C H E M O S TAT Prep 1 Prep 2...
  • Page 2 Summary of Clinical Data: Clinical data summary information that was, in part, the basis for granting the de novo can be found on the Cook Medical website at CookMedical.com/HemosprayData Prior to removing Hemospray device from Upon completion of procedure, depressurize patient, turn valve to closed position.