8. HYGIENE TREATMENT
82 •
seca mVSA 535
➔
Cleaning
➔
Disinfecting
➔
Sterilizing
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
► Ensure that the device is switched off before performing any hygiene
treatment.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
► Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
► Use disinfectants including the active ingredient 70 % isopropanol
only for measuring accessories for measuring vital signs.
➔
Removing/fitting probe holder (devices
with temperature probe)