Repair and Replacement Procedures

Removing and Replacing the A5 through A10 Boards

Removing and Replacing the A5 through A10 Boards
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-6
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. With a T-10 TORX driver, remove the two brackets that secure cables to the top of the
midweb.
4. Identify the board you want to remove and disconnect any cables that are attached to it.
5. Lift the two extractors (item
Before removing the board completely, check the bottom of the board for any
NOTE
attached cables.
6. While holding on to the extractors, slide the board out of the slot and remove it from the
analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-16
"Removing the Covers" on page
), located at each end of the board.
Table 7-2 on page
PNA Series Microwave Network Analyzers
7-45.
Service Guide N5230-90014
N5230A
7-6.