Er 81 Generator processor EEPROM failure
Er 82 Keyboard processor configuration error
Er 83 Generator processor configuration error
Er 84 Tube power generator time out

3.11 Errors 90-99 (other errors)

ER 90 Keyboard processor
ER 91Keyboard processor
Er 99 Error code generation error
Technical Manual
- Replace the microprocessor chip or the complete board.
There is some non volatile memory for various parameters in the microprocessor chip itself. The
writing of this memory is limited to a minimum of 10 000 times that is in practice 20 years of
equipment life.
However, if this memory did fail, this error is indicated.
- Replace the microprocessor chip or the complete board.
The keyboard processor chip watchdog circuit is not activated.
-Change the keyboard processor board
The generator processor watchdog circuit is not activated.
- Change the generator processor board
Exposure backup timer: if the exposure time exceeds 40 s the generator is stopped with this error
indication. If this error appears more than once, replace the Generator processor board.
Keyboard processor EEPROM not activated.
- Change keyboard processor board
Keyboard processor stack overflow.
- Change the software on the keyboard processor board.
This error indication can only be result of non matching software versions on generator and key-
board processors or an indication of defective keyboard processor board.
- Check program version numbers and/or replace keyboard processor