Planmeca Proline CC Technical Manual page 101

With pan/ceph/dimax3
Table of Contents

Advertisement

Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
By repeatedly pressing this key you can select each slot in turn so that they can be checked and,
if necessary, adjusted.
UP
UP
Primary slot number 2
Primary slot number 3
DOWN
DOWN
Primary slot number 2
Primary slot number 3
Figure 18
When all the slots have been checked, and if necessary adjusted, exit the service mode and take
a ball phantom exposure to check the position of the patient positioning mechanism, see section
"PATIENT POSITIONING MECHANISM" on page E-27.
Technical Manual
Planmeca Proline CC panoramic x-ray E-13