6-30 | Model 241CE II Hydrocarbon Dewpoint Analyzer
5. If a lubricant was used to help dislodge the Window Housing during
removal, clean the lubricant off the Analyzer Housing before replacing
the Window Housing.
Align the push-button holes in the Window Housing (original part,
if it is not damaged, or new part if the original part is being replaced)
with the push-buttons and carefully place it over the buttons, being
careful not to damage the housing or buttons.
Secure it to the analyzer housing using the (12) M6 x 16 screws. Tight-
en the screws in a cross-pattern. The torque specification to fasten M6
screws is 4.0 Nm, ±0.7 Nm (36 in.-lb, ±9 in.-lb).

Window (User Interface)

Periodically, examine the window for scratches, chips, or cracks. Replace
the Window Housing assembly if it is damaged. Replace the push-buttons
at the same time.
!
Be careful not to scratch the window. Scratches will reduce impact
resistance.
CAUTION