Working In The Configuration Menu - Ametek 241CE II Operator's Manual

Hydrocarbon dewpoint analyzer
Hide thumbs Also See for 241CE II:
Table of Contents

Advertisement

A password is required
to access the MonitorCFG
menu.
Sub-Menu Item
Passcode2?
Water Measurement Option
4-18 | Model 241CE II Hydrocarbon Dewpoint Analyzer

Working in the Configuration Menu

The Configuration (MonitorCFG) menu allows you to customize unit-
specific parameters.
To view the MonitorCFG menu, press  until MonitorCFG appears. To
view the sub-menu items within this menu, press . Enter the password
at the "Passcode2?" prompt to view the MonitorCFG menu.
The MonitorCFG sub-menu items (listed in their viewing order) and their
descriptions are:
IMPORTANT
Do not change parameters marked '*' (under "Description") unless
NOTE
instructed to do so by an AMETEK service representative. Doing so
will result in incorrect measurements and readings.
Description
This prompt asks to enter Passcode2 to access the MonitorCFG menu (the factory default password
is '- - - -'). Enter this password to view and/or change the parameters in this menu.
The Water Measurement Option sub-menu allows you to select the type of information you wish
to display. Only one of the options can be selected at any time. Note that "Water Measurement
Option" is not displayed; only the current setting will be displayed (e.g., H2O DP:EXT). The
available options that can be selected include:
• H2O DP:EXT (Water Dewpoint, external measurement – requires external water signal).
• H2O CT:EXT (Water Content, external measurement – requires external water signal).
• H2O DP:INT (Water Dewpoint, internal measurement).
• H2O DP:DISAB (Water Dewpoint, disabled).
Note: If you change the Water Measurement Option parameter, you will also have
to configure the analyzer to display the associated units of measurement
from the Options code sub-menu, depending on the information you want to
be analyzed and displayed. These changes are required to ensure accurate
representation of the output signals you require. Details about these changes
are described below.
If changes are made to the Water Measurement Option parameter, you will also have to change the
Options code parameters:
• If H2O DP:EXT is enabled, select an option from the Options code sub-menu:
- DP&m (dewpoint, metric units).
- DP&i (dewpoint, imperial units).
• If H2O CT:EXT is enabled, select an option from the Options code sub-menu:
- WC&m (water content, metric units).
- WC&i (water content, imperial units).
- PMVm (parts per million by volume, metric units).
- PMVi (parts per million by volume, imperial units).
- PMWm (parts per million by weight, metric units).
- PMWi (parts per million by weight, imperial units).
(Continued next page)