Environmental
Cooling Capability
Minimum Clearances
2-4 | Model 241CE II Hydrocarbon Dewpoint Analyzer
IEC Installation Category II
IEC Pollution Degree 2
Humidity
0–95 % RH (non-condensing)
Maximum Altitude
2000 metres (6560 feet)
Typically 60 °C (108 °F) below the surrounding ambient temperature at the
location of the analyzer. The cooling capability of the analyzer is affected
by the sample gas composition and pressure.
60 °C is achieved at a 20 °C ambient temperature. The achievable tem-
perature difference will be slightly higher as the ambient temperature
NOTE
increases above 20 °C.
Heat Dissipation
To allow for heat dissipation and servicing, the required minimum
clearances are:
- 150 mm (6") each side
- 75 mm (3") top and bottom
- 900 mm (36") front
Door Swing
To allow for full door swing, minimum clearance in front of the analyzer is
approximately 574 mm (22.6").