Dry Gas Supply (Purge Air)
A dry atmosphere within the Thermoelectric Cooler (TEC) compartment
is necessary to prevent ice from forming, which can damage the cooler. A
dry atmosphere can be created by connecting the Flame Arrestor Fitting
on the Cooler Housing to a purge gas supply (typically N
supply – such as air – with a water dewpoint temperature less than -40 °C
(-40 °F) and a flow rate less than 15 L/day (0.53 CFD). The purge gas pres-
sure upstream of the Flame Arrestor Fitting cannot exceed atmospheric
pressure by more than 1 PSIG (70 mBAR). (Dry gas supplied by the cus-
tomer/end user.)
!
CAUTION
Approvals and Certifications
UL/CSA:
ATEX:
IECEx :
GOST:
CE Compliance:
EMC:
LVD:
PED:
RoHS:
Before operating the Model 241CE II Analyzer, a Flame Arrestor Fit-
ting must be installed on the Cooler Housing (to a dry gas supply).
This fitting and an associated Flow Restrictor is supplied by AM-
ETEK. Failure to install a dry gas supply meeting these specifications
voids the warranty of the Peltier Cooler.
CSA General Safety Requirements
Class I, Division 1, Groups C and D; Ex d IIB T3 (Class I,
Zone 1, Group IIB)
Certificate No.:
Certificate No.:
1ExdIIBT3 X
GOST Pattern Approval
Complies with all relevant European Directives
Electromagnetic Compatibility Directive 2014/30/EU,
with EN 61326 Industrial
Low Voltage Directive 2014/35/EU, with EN 61010-1
Pressure Equipment Directive 2014/68/EU
Article 4, Paragraph 3
Restriction of Hazardous Substances Directive 2011/65/EU
(RoHS 2)
) or to a dry gas
2
KEMA 02ATEX2121 X
II 2 G Ex db IIB T4 Gb
IECEx DEK12.0052X
Ex db IIB T4 Gb
Specifications
|
2-5